Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 20 de 18.747
Filtrar
1.
Artigo em Inglês | MEDLINE | ID: mdl-38618610

RESUMO

Left ventricular assist devices serve as a salvage therapy for patients with advanced heart failure. Complications such as thrombosis and obstruction can lead to acute device malfunction, posing significant clinical risks. A multidisciplinary approach is crucial for management. Few cases in the literature have demonstrated the safety and efficacy of percutaneous intervention, which holds significant value due to its less invasive nature and minimal risk of morbidity, especially in high-risk surgical patients.


Assuntos
Insuficiência Cardíaca , Coração Auxiliar , Trombose , Humanos , Trombose/diagnóstico por imagem , Trombose/etiologia , Trombose/terapia , Insuficiência Cardíaca/diagnóstico , Insuficiência Cardíaca/etiologia , Insuficiência Cardíaca/terapia
2.
Ann Plast Surg ; 92(4S Suppl 2): S200-S203, 2024 Apr 01.
Artigo em Inglês | MEDLINE | ID: mdl-38556673

RESUMO

BACKGROUND: The use of left ventricular assist devices (LVADs) for patients with end-stage cardiac failure awaiting heart transplantation has become increasingly common. However, ventricular assist device-related infections remain a major problem complicating their long-term use. Retrospective review has previously shown a decrease in lifetime return to operating room (RTOR) with no change in 90-day complications when a muscle or omental flap is used for coverage after washout for infection. We wished to determine if early plastic surgery intervention led to a decreased length of stay for these patients. METHODS: Patients with LVAD readmitted for LVAD infection at a single institution from 2008 to 2021 were identified using a preexisting database. Patients were followed retrospectively for an average of 3.2 years. Patient demographics, preoperative diagnosis/disease state, type of ventricular assist device inserted, postoperative day of ventricular assist device infection onset, definitive device coverage, timing of coverage procedure after the initial washout for infection, type of flap used for coverage, 90-day complications after definitive coverage, and lifetime return to operating room for infection were reviewed. Comparison analysis with χ2 and analysis of variance testing was used to analyze outcomes. RESULTS: Of 568 patients admitted with an LVAD infection, 104 underwent operative debridement and closure by plastic and reconstructive surgery (PRS) or cardiothoracic surgery (CTS). Fifty-three underwent PRS closure, and 51 underwent CTS closure. There was an increased incidence of diabetes among the PRS group (P < 0.001); otherwise, there was no difference in baseline characteristics. There was increased RTOR over a lifetime with CTS closure compared with PRS (P = 0.03) and increased 90-day risk of infection (P = 0.007). Patients with PRS closure had an increased risk of postoperative hematoma (P = 046). Plastic and reconstructive surgery was typically consulted on hospital day 10. Both PRS and CTS closure patients were discharged on postoperative day 7, on average (P = 0.542). CONCLUSIONS: Plastic surgery involvement with surgical decision making and closure of infected LVAD devices has a decrease in lifetime RTOR and decreased 90-day complications related to infections. There may be a benefit to earlier PRS consultation for coverage assessment.


Assuntos
Insuficiência Cardíaca , Coração Auxiliar , Infecções Relacionadas à Prótese , Cirurgia Plástica , Humanos , Coração Auxiliar/efeitos adversos , Estudos Retrospectivos , Salas Cirúrgicas , Infecções Relacionadas à Prótese/etiologia , Insuficiência Cardíaca/cirurgia , Insuficiência Cardíaca/etiologia , Resultado do Tratamento
3.
Eur Rev Med Pharmacol Sci ; 28(6): 2199-2206, 2024 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-38567583

RESUMO

OBJECTIVE: Pediatric heart failure is an important cause of morbidity and mortality in childhood. Left ventricular assist devices (L-VAD) are used for bridging to transplantation in patients with indications for heart transplantation. PATIENTS AND METHODS: The children included in the study were patients who underwent implantation of an L-VAD due to advanced heart failure at Ege University Faculty of Medicine Hospital between January 2009 and January 2023. RESULTS: Of the 33 patients who underwent L-VAD implantation, 16 (48.5%) were female and 17 (51.5%) were male. The median age at surgery was 13 years (IQR, 9.5-15). The median weight was 44 kg (IQR, 25.65-52), the median height was 158 cm (IQR, 134.5-168.5), and the median body surface area was 1.37 m2 (IQR, 0.95-1.51). All patients who underwent L-VAD implantation had an echocardiographic diagnosis of dilated cardiomyopathy. The patients underwent a median of 16 (IQR, 9-21) ECGs, and the median number of 24-hour Holter ECGs obtained was 3 (IQR, 2-5). Arrhythmias that occurred after L-VAD implantation were classified as atrial and ventricular. Ventricular arrhythmia included ventricular tachycardia (VT) lasting for more than 30 seconds (sustained VT), VT lasting for less than 30 seconds (nonsustained VT), and ventricular fibrillation. Atrial arrhythmias included atrial flutter, atrial fibrillation, supraventricular tachycardia, and atrial ectopic tachycardia. During the follow-up, atrial or ventricular arrhythmias were observed in 11 (33%) patients. The most common rhythm disturbances before L-VAD implantation were ventricular arrhythmias, while after the surgery, atrial arrhythmias were found to be the most frequent. A total of 5 patients underwent cardioversion (n=2) or defibrillation (n=3) due to arrhythmia. CONCLUSIONS: In patients undergoing L-VAD implantation, rhythm disorders that could normally lead to hemodynamic instability are frequently encountered. In these rhythm disorders, medical therapy should be attempted before resorting to cardioversion or defibrillation, and subsequently, more aggressive treatment methods should be considered.


Assuntos
Fibrilação Atrial , Insuficiência Cardíaca , Coração Auxiliar , Taquicardia Supraventricular , Taquicardia Ventricular , Humanos , Masculino , Feminino , Criança , Adolescente , Coração Auxiliar/efeitos adversos , Insuficiência Cardíaca/terapia , Insuficiência Cardíaca/complicações
4.
J Cardiothorac Surg ; 19(1): 167, 2024 Apr 01.
Artigo em Inglês | MEDLINE | ID: mdl-38561765

RESUMO

BACKGROUND: The implantation of left ventricular assist devices (LVADs) as a bridge to transplantation or as destination therapy in end-stage heart failure patients is frequently complicated by the emergence of ventricular arrhythmias (VAs). These arrhythmias have been implicated in precipitating deleterious clinical outcomes, increased mortality rates and augmented healthcare expenditures. CASE PRESENTATION: We present a challenging case of a 49-year-old male with a history of dilated cardiomyopathy who received an LVAD. Post-implantation, the patient suffered from intractable VAs, leading to multiple rehospitalizations and hemodynamic deterioration. Despite exhaustive medical management and electrical cardioversion attempts, the patient's VAs persisted, ultimately necessitating prioritization for cardiac transplantation. DISCUSSION: This case highlights the challenges in managing VAs in LVAD patients and the importance of multidisciplinary collaboration. While pharmacological intervention is the initial strategy, catheter ablation may be considered in selected cases when medication is insufficient. In instances of intractable VAs, expeditious listing for heart transplantation as a high-priority candidate is advisable when feasible.


Assuntos
Insuficiência Cardíaca , Coração Auxiliar , Masculino , Humanos , Pessoa de Meia-Idade , Coração Auxiliar/efeitos adversos , Insuficiência Cardíaca/complicações , Insuficiência Cardíaca/cirurgia , Arritmias Cardíacas/etiologia , Hemodinâmica , Cardioversão Elétrica , Resultado do Tratamento
6.
N Engl J Med ; 390(15): 1382-1393, 2024 Apr 18.
Artigo em Inglês | MEDLINE | ID: mdl-38587239

RESUMO

BACKGROUND: The effects of temporary mechanical circulatory support with a microaxial flow pump on mortality among patients with ST-segment elevation myocardial infarction (STEMI) complicated by cardiogenic shock remains unclear. METHODS: In an international, multicenter, randomized trial, we assigned patients with STEMI and cardiogenic shock to receive a microaxial flow pump (Impella CP) plus standard care or standard care alone. The primary end point was death from any cause at 180 days. A composite safety end point was severe bleeding, limb ischemia, hemolysis, device failure, or worsening aortic regurgitation. RESULTS: A total of 360 patients underwent randomization, of whom 355 were included in the final analysis (179 in the microaxial-flow-pump group and 176 in the standard-care group). The median age of the patients was 67 years, and 79.2% were men. Death from any cause occurred in 82 of 179 patients (45.8%) in the microaxial-flow-pump group and in 103 of 176 patients (58.5%) in the standard-care group (hazard ratio, 0.74; 95% confidence interval [CI], 0.55 to 0.99; P = 0.04). A composite safety end-point event occurred in 43 patients (24.0%) in the microaxial-flow-pump group and in 11 (6.2%) in the standard-care group (relative risk, 4.74; 95% CI, 2.36 to 9.55). Renal-replacement therapy was administered to 75 patients (41.9%) in the microaxial-flow-pump group and to 47 patients (26.7%) in the standard-care group (relative risk, 1.98; 95% CI, 1.27 to 3.09). CONCLUSIONS: The routine use of a microaxial flow pump with standard care in the treatment of patients with STEMI-related cardiogenic shock led to a lower risk of death from any cause at 180 days than standard care alone. The incidence of a composite of adverse events was higher with the use of the microaxial flow pump. (Funded by the Danish Heart Foundation and Abiomed; DanGer Shock ClinicalTrials.gov number, NCT01633502.).


Assuntos
Coração Auxiliar , Infarto do Miocárdio com Supradesnível do Segmento ST , Choque Cardiogênico , Idoso , Feminino , Humanos , Masculino , Coração Auxiliar/efeitos adversos , Incidência , Choque Cardiogênico/etiologia , Choque Cardiogênico/mortalidade , Choque Cardiogênico/cirurgia , Infarto do Miocárdio com Supradesnível do Segmento ST/complicações , Infarto do Miocárdio com Supradesnível do Segmento ST/mortalidade , Infarto do Miocárdio com Supradesnível do Segmento ST/terapia , Resultado do Tratamento , Circulação Assistida/efeitos adversos , Circulação Assistida/instrumentação , Circulação Assistida/métodos
8.
Clin Transplant ; 38(4): e15317, 2024 04.
Artigo em Inglês | MEDLINE | ID: mdl-38607287

RESUMO

BACKGROUND: United Network for Organ Sharing (UNOS) allocation criteria changed in 2018 to accommodate the increased prevalence of patients on a ventricular assist device as a bridge to heart transplant and prioritize sicker people in anticipation of a heart graft. We aimed to assess the impact of patient age in the new allocation policy on mortality following heart transplantation. Secondary outcomes included the effect of age ≥70 on post-transplant events, including stroke, dialysis, pacemaker, and rejection requiring treatment. METHODS: The UNOS Registry was queried to identify patients who underwent heart transplants alone in the US between 2000 and 2021. Patients were divided into groups according to their age (over 70 and under 70 years old). RESULTS: Patients aged over 70 were more likely to require dialysis during follow-up, but less likely to experience rejection requiring treatment, compared with patients aged <70. Age ≥70 in the new allocation system was a significant predictor of 1-year mortality (adjusted HR: 1.41; 95% CI: 1.05-1.91; p = .024), but its effect on 5-year mortality was not significant after adjusting for potential confounders (adjusted HR: 1.27; 95% CI:.97-1.66; p = .077). Undergoing transplantation under the new allocation policy vs the old allocation policy was not a significant predictor of mortality in patients over 70 years old. CONCLUSIONS: Age ≥70 is a significant predictor of 1-year mortality following heart transplantation, but not at 5 and 10 years; however, the new allocation does not seem to have changed the outcomes for this group of patients.


Assuntos
Transplante de Coração , Coração Auxiliar , Marca-Passo Artificial , Humanos , Idoso , Idoso de 80 Anos ou mais , Sistema de Registros , Diálise Renal
9.
Braz J Cardiovasc Surg ; 39(3): e20230258, 2024 Apr 17.
Artigo em Inglês | MEDLINE | ID: mdl-38630058

RESUMO

INTRODUCTION: Limited options in the end-stage treatment of heart failure have led to increased use of left ventricular assist devices. For this reason, the rate of non-cardiac surgeries in patients with left ventricular assist devices is also increasing. Our study aims to analyze surgical rate, anesthesia management, and results by reviewing our 11-year experience with patients who underwent non-cardiac surgery receiving left ventricular assist devices support. METHODS: We retrospectively evaluated 57 patients who underwent non-cardiac surgery and 67 non-cardiac surgical procedures among 274 patients who applied between January 2011 and December 2022 and underwent left ventricular assist devices implantation with end-stage heart failure. RESULTS: Fifty (74.6%) patients with left ventricular assist devices admitted to the hospital for non-cardiac surgery were emergency interventions. The most common reasons for admission were general surgery (52.2%), driveline wound revision (22.3%), and neurological surgery (14.9%). This patient group has the highest in-hospital mortality rate (12.8%) and the highest rate of neurological surgery (8.7%). While 70% of the patients who underwent neurosurgery were taken to surgery urgently, the International Normalized Ratio values of these patients were between 3.5 and 4.5 at the time of admission to the emergency department. CONCLUSION: With a perioperative multidisciplinary approach, higher morbidity and mortality risks can be reduced during emergencies and major surgical procedures.


Assuntos
Insuficiência Cardíaca , Coração Auxiliar , Humanos , Estudos Retrospectivos , Hospitais , Ventrículos do Coração , Insuficiência Cardíaca/cirurgia
10.
Cardiol Clin ; 42(2): 187-193, 2024 May.
Artigo em Inglês | MEDLINE | ID: mdl-38631789

RESUMO

Cardiogenic shock is a lethal condition with significant morbidity, characterized by myocardial insults leading to low cardiac output and ensuing systemic hypoperfusion. While mortality rates remain high, we have improved upon our recognition and definition of cardiogenic shock, now with an emphasis on defining stages of shock to help guide effective treatment strategies with either pharmacologic or mechanical circulatory support. In this review, the authors summarize these stages as well as discuss indications, function, selection, and troubleshooting of the various temporary mechanical circulatory support devices.


Assuntos
Coração Auxiliar , Choque Cardiogênico , Humanos , Choque Cardiogênico/terapia , Resultado do Tratamento , Balão Intra-Aórtico
11.
Eur J Cardiothorac Surg ; 65(4)2024 Mar 29.
Artigo em Inglês | MEDLINE | ID: mdl-38613845

RESUMO

OBJECTIVES: Myocardial recovery in children supported by a durable left ventricular assist device is a rare, but highly desirable outcome because it could potentially eliminate the need for a cardiac transplant and the lifelong need for immunosuppressant therapy and the risk of complications. However, experience with this specific outcome is extremely limited. METHODS: All patients < 19 years old supported by a durable left ventricular assist device from the European Registry for Patients with Mechanical Circulatory Support database were included. Participating centres were approached for additional follow-up data after explantation. Associated factors for explantation due to myocardial recovery were explored using Cox proportional hazard models. RESULTS: The incidence of recovery in children supported by a durable left ventricular assist device was 11.7% (52/445; median duration of support, 122.0 days). Multivariable analyses showed body surface area (hazard ratio 0.229; confidence interval 0.093-0.565; P = 0.001) and a primary diagnosis of myocarditis (hazard ratio 4.597; confidence interval 2.545-8.303; P < 0.001) to be associated with recovery. Left ventricular end-diastolic diameter in children with myocarditis was not associated with recovery. Follow-up after recovery was obtained for 46 patients (88.5%). Sustained myocardial recovery was reported in 33/46 (71.7%) at the end of the follow-up period (28/33; >2 year). Transplants were performed in 6/46 (11.4%) (in 5 after a ventricular assist device was reimplanted). Death occurred in 7/46 (15.2%). CONCLUSIONS: Myocardial recovery occurs in a substantial portion of paediatric patients supported with durable left ventricular assist devices, and sustainable recovery is seen in around three-quarters of them. Even children with severely dilated ventricles due to myocarditis can show recovery. Clinicians should be attentive to (developing) myocardial recovery. These results can be used to develop internationally approved paediatric weaning guidelines.


Assuntos
Insuficiência Cardíaca , Transplante de Coração , Coração Auxiliar , Miocardite , Humanos , Criança , Adulto Jovem , Adulto , Coração Auxiliar/efeitos adversos , Miocardite/cirurgia , Miocárdio , Diástole , Insuficiência Cardíaca/cirurgia , Resultado do Tratamento
12.
Intensive Crit Care Nurs ; 82: 103658, 2024 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-38431985

RESUMO

BACKGROUND: Surgical wound infection is the most frequent type of care health associated infection. Lack of knowledge about the prevention of surgical wound infection in patients undergoing left ventricular assist device implantation could significantly undermine the potential benefits of surgical intervention. OBJECTIVES: This study aimed to map the recommendations for adult patients undergoing left ventricular assist device implantation. DESIGN: This is a scoping review, being registered in the Open Science Framework under DOI https://doi.org/10.17605/OSF.IO/Q76B3 (https://osf.io/q76b3/). METHOD: Left ventricular assist device coordinators and nurse specialists in dermatology and stomatherapy conducted a scoping review in Scopus, The Cochrane Database of Systematic Reviews, Cumulative Index to Nursing and Allied Health Literature (CINAHL), limited to the period between 2015 and 2022. The results of this scoping review will be discussed and presented in separate articles. This paper will synthesize research evidence on the perioperative topic. RESULTS: The initial searches resulted in 771 studies. Sixty nine met the eligibility criteria and were included in the scoping review. Eight articles addressing the perioperative topic that answered the question of this article were included. CONCLUSION: Although this scoping review included heterogeneous, and scarce studies with left ventricular assist device patients. As such, there are many promising future research directions for this topic. IMPLICATIONS FOR CLINICAL PRACTICE: Infection surveillance should be an integral part of left ventricular assist device implantation programs in health care institutions. Velvet completely buried in subcutaneous tissues reduces transmission system infection. Triple tunnel method reduces transmission system infection risk.


Assuntos
Coração Auxiliar , Infecção da Ferida Cirúrgica , Adulto , Humanos , Infecção da Ferida Cirúrgica/prevenção & controle , Coração Auxiliar/efeitos adversos , Revisões Sistemáticas como Assunto
13.
Intensive Care Med ; 50(4): 493-501, 2024 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-38526578

RESUMO

An implantable left ventricular assist device (LVAD) is indicated as a bridge to transplantation or recovery in the United Kingdom (UK). The mechanism of action of the LVAD results in a unique state of haemodynamic stability with diminished arterial pulsatility. The clinical assessment of an LVAD recipient can be challenging because non-invasive blood pressure, pulse and oxygen saturation measurements may be hard to obtain. As a result of this unusual situation and complex interplay between the device and the native circulation, resuscitation of LVAD recipients requires bespoke guidelines. Through collaboration with key UK stakeholders, we assessed the current evidence base and developed guidelines for the recognition of clinical deterioration, inadequate circulation and time-critical interventions. Such guidelines, intended for use in transplant centres, are designed to be deployed by those providing immediate care of LVAD patients under conditions of precipitous clinical deterioration. In summary, the Joint British Societies and Transplant Centres LVAD Working Group present the UK guideline on management of emergencies in implantable LVAD recipients for use in advanced heart failure centres. These recommendations have been made with a UK resuscitation focus but are widely applicable to professionals regularly managing patients with implantable LVADs.


Assuntos
Deterioração Clínica , Insuficiência Cardíaca , Transplante de Coração , Coração Auxiliar , Humanos , Emergências , Insuficiência Cardíaca/terapia
14.
Med Eng Phys ; 125: 104124, 2024 03.
Artigo em Inglês | MEDLINE | ID: mdl-38508801

RESUMO

BACKGROUND: The frequent occurrence of thromboembolic cerebral events continues to limit the widespread implementation of Ventricular Assist Devices (VAD) despite continued advancements in VAD design and anti-coagulation treatments. Recent studies point to the optimal positioning of the outflow graft (OG) as a potential mitigator of post implantation thromboembolism. OBJECTIVE: This study aims to examine the tailoring of the OG implantation orientation with the goal of minimizing the number of thrombi reaching the cerebral vessels by means of a formal shape optimization scheme incorporated into a multi-scale hemodynamics analysis. METHODS: A 3-D patient-specific computational fluid dynamics model is loosely coupled in a two-way manner to a 0-D lumped parameter model of the peripheral circulation. A Lagrangian particle-tracking scheme models and tracks thrombi as non-interacting solid spheres. The loose coupling between CFD and LPM is integrated into a geometric shape optimization scheme which aims to optimize an objective function that targets a drop in cerebral embolization, and an overall reduction in particle residence times. RESULTS: The results elucidate the importance of OG anastomosis orientation and placement particularly in the case that studied particle release from the OG, as a fivefold decrease in cerebral embolization was observed between the optimal and non-optimal implantations. Another case considered particle release from the ventricle and aortic root walls, in which optimal implantation was achieved with a shallow insertion angle. Particle release from all three origins was investigated in the third case, demonstrating that the optimal configurations were generally characterized by VAD flow directed along the central lumen of the aortic arch. Because optimal configurations depended on the anatomic origin of the thrombus, it is important to determine, in clinical studies, the most likely sites of thrombus formation in VAD patients.


Assuntos
Coração Auxiliar , Trombose , Humanos , Cânula , Aorta , Aorta Torácica , Hemodinâmica , Modelos Cardiovasculares
15.
Sci Rep ; 14(1): 7443, 2024 03 28.
Artigo em Inglês | MEDLINE | ID: mdl-38548818

RESUMO

With improved treatment of coronary artery disease, more patients are surviving until heart failure occurs. This leads to an increase in patients needing devices for struggling with heart failure. Ventricular assist devices are known as the mainstay of these devices. This study aimed to design a centrifugal pump as a ventricular assist device. In order to design the pump, firstly, the geometrical parameters of the pump, including the gap distance, blade height, and position of the outlet relative to the blade, were investigated. Finally, the selected configuration, which had all the appropriate characteristics, both hydraulically and physiologically, was used for the rest of the study. The study of the blade, as the main component in energy transfer to the blood, in a centrifugal pump, has been considered in the present study. In this regard, the point-to-point design method, which is used in industrial applications, was implemented. The designer chooses the relationship between the blade angles at each radius in the point-to-point method. The present study selected logarithmic and second-order relations for designing the blade's profile. In total, 58 blades were examined in this study, which differed regarding blade inlet and outlet angles and the relationship between angle and radial position. ANSYS CFX 17.0 software was utilized to simulate blades' performances, and a benchmark pump provided by the US Food and Drug Administration (FDA) was used to validate the numerical simulations. Then, the selected impeller from the numerical investigation was manufactured, and its performance was compared experimentally with the FDA benchmark pump. A hydraulic test rig was also developed for experimental studies. The results showed that among the blades designed in this study, the blade with an input angle of 45° and an output angle of 55°, which is designed to implement a logarithmic relationship, has the best performance. The selected impeller configuration can increase the total head (at least by 20%) at different flow rates compared to the FDA pump.


Assuntos
Insuficiência Cardíaca , Coração Auxiliar , Humanos , Desenho de Prótese , Desenho de Equipamento
16.
Int Heart J ; 65(2): 218-229, 2024.
Artigo em Inglês | MEDLINE | ID: mdl-38556333

RESUMO

Troponin (Tn) is a biomarker related to myocardial necrosis and is elevated in patients with myocarditis. This study aimed to investigate the association between cardiac Tn levels and the course of cardiac function, and prognosis in patients with fulminant myocarditis (FM) receiving percutaneous mechanical circulatory support (MCS).We used data from a multicenter retrospective registry, CHANGE PUMP 2, which included 216 patients with FM who required MCS. Among them, 141 patients whose Tn levels were available were analyzed. The patients were divided into low and high Tn groups according to the median values of TnT and TnI.The median age was 54 years, and 59.6% were male. The TnT and TnI on day 1 (at MCS initiation) were 3.8 (1.4-10.0) and 21.4 (8.4-68.8) ng/mL. While the left ventricular ejection fraction (LVEF) was similar on day 1 (25.0% versus 24.5%), the low Tn group showed better LVEF improvement on day 7 than the high Tn group (45.0% versus 25.3%, P < 0.001). LVEF at 1 year after admission was higher in the low Tn group (65.0% versus 59.7%, P = 0.039). The low Tn group had a better 90-day composite endpoint in death, durable left ventricular assist device implantation, and heart transplantation compared to the high Tn group (hazard ratio 0.47, 95% CI 0.23-0.95).Tn levels were associated with short- and long-term cardiac recovery and adverse outcomes in patients with FM receiving MCS due to cardiogenic shock.


Assuntos
Coração Auxiliar , Miocardite , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Miocardite/diagnóstico , Prognóstico , Estudos Retrospectivos , Choque Cardiogênico , Volume Sistólico , Resultado do Tratamento , Troponina , Função Ventricular Esquerda , Estudos Multicêntricos como Assunto
17.
J Card Fail ; 30(4): 596-609, 2024 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-38431185

RESUMO

The field of durable mechanical circulatory support (MCS) has undergone an incredible evolution over the past few decades, resulting in significant improvements in longevity and quality of life for patients with advanced heart failure. Despite these successes, substantial opportunities for further improvements remain, including in pump design and ancillary technology, perioperative and postoperative management, and the overall patient experience. Ideally, durable MCS devices would be fully implantable, automatically controlled, and minimize the need for anticoagulation. Reliable and long-term total artificial hearts for biventricular support would be available; and surgical, perioperative, and postoperative management would be informed by the individual patient phenotype along with computational simulations. In this review, we summarize emerging technological innovations in these areas, focusing primarily on innovations in late preclinical or early clinical phases of study. We highlight important considerations that the MCS community of clinicians, engineers, industry partners, and venture capital investors should consider to sustain the evolution of the field.


Assuntos
Insuficiência Cardíaca , Coração Artificial , Coração Auxiliar , Humanos , Insuficiência Cardíaca/cirurgia , Invenções , Qualidade de Vida
18.
Catheter Cardiovasc Interv ; 103(5): 799-802, 2024 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-38461378

RESUMO

Ramp studies are utilized for speed optimization of continuous flow left ventricular assist devices (CF-LVADs). We here report the utility of combined left and right heart catheterization during a ramp study to ensure a comprehensive understanding of the hemodynamic implications on both ventricles. Pressure-volume loop (PV loop) monitoring uncovered compromised systolic and mildly compromised right ventricular function with increasing LVAD speeds, despite improvement in left ventricular unloading. These findings informed patient management and highlight the potential utility of PV loop monitoring as an adjunct to left and right heart catheterization during ramp studies of next-generation LVADs.


Assuntos
Insuficiência Cardíaca , Coração Auxiliar , Humanos , Função Ventricular Direita , Resultado do Tratamento , Hemodinâmica , Cateterismo Cardíaco , Insuficiência Cardíaca/diagnóstico , Insuficiência Cardíaca/terapia , Função Ventricular Esquerda
19.
Int J Artif Organs ; 47(3): 162-172, 2024 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-38450429

RESUMO

BACKGROUND: The hydrodynamic suspension structure design of the axial blood pump impeller can avoid the problems associated with using mechanical bearings. However, the particular impeller structure will impact the hydraulic performance and hemolysis of the blood pump. METHOD: This article combines computational fluid dynamics (CFD) with the Lagrange particle tracking method, aiming to improve the blood pump's hydraulic and hemolysis performance. It analyzes the flow characteristics and hemolysis performance inside the pump. It optimizes the taper of the impeller hub, the number of blades, and the inclination angle of the circumferential groove at the top of the blade. RESULTS: Under certain rotational speed conditions, an increase in the taper of the impeller hub or the number of blades can increase the pumping pressure of a blood pump, but an increase in the number of blades will reduce the flow rate. The design of circumferential grooves at the top of the blade can increase the pumping pressure to a certain extent, with little impact on the hemolysis performance. The impeller structure is optimized based on the estimated hemolysis of each impeller model blood pump. It could be seen that when the pump blood pressure and flow rate were reached, the optimized impeller speed was reduced by 11.4%, and the estimated hemolysis value was reduced by 10.5%. CONCLUSION: In this paper, the rotor impeller structure of the blood pump was optimized to improve the hydraulic and hemolytic performance effectively, which can provide a reference for the related research of the axial flow blood pump using hydraulic suspension.


Assuntos
Coração Auxiliar , Humanos , Desenho de Equipamento , Hemólise , Simulação por Computador , Pressão Sanguínea
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...